Description
Dräger Perseus A500 software versions 2.00 through 2.02 contains an improper input handling vulnerability that allows external attackers to cause a denial of service by sending specifically crafted non-Medibus-compliant data through the Medibus interface. Attackers can overload the internal processor with malformed data to trigger a warm restart, causing ventilation pressure to drop to ambient level and interrupting ventilation for several seconds before therapy resumes.
References2
Metrics
- CVSS 3.1
-
4.0
MEDIUM
Score source: CNA
CVSS:3.1/AV:N/AC:H/PR:N/UI:N/S:C/C:N/I:N/A:L - CVSS 4.0
-
6.3
MEDIUM
Score source: CNA
CVSS:4.0/AV:N/AC:H/AT:N/PR:N/UI:N/VC:N/VI:N/VA:L/SC:N/SI:N/SA:L/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X - EPSS — exploit probability
- 0% percentile: 15 · as of 15 November 2026
- CWE
- CWE-1286
Could this be exploited in your environment?
A patch closes the known hole. A penetration test shows what an attacker can still reach — with proof.